EasyManuals Logo
Home>Agilent Technologies>Test Equipment>E4406A VSA Series

Agilent Technologies E4406A VSA Series Service Guide

Agilent Technologies E4406A VSA Series
266 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #252 background imageLoading...
Page #252 background image
252 Chapter 5
Post-Repair Procedures
Internal 50 MHz Calibrator Adjustment
used between the power sensor and the reference of the power meter.
The 0 dBm 50 MHz power reference from the power meter could
damage the highly sensitive power sensor.
2. Zero and calibrate the power meter.
3. Configure the source, power meter, and power sensor as shown in
Figure 5-2.
4. Preset the signal source and initialize the test equipment
parameters as listed in Table 5-4, “Signal Source Initialization,” on
page 252
5. Adjust the source amplitude until the power meter reads
25 dBm ±±0.2 dBm.
6. Record the measured value from the power meter.
___________ dBm
NOTE Do not readjust the source amplitude during the remainder of the test.
7. Configure equipment as shown in Figure 5-3, “Internal 50 MHz
Calibration Adjustment Setup”.
8. Press
Preset.
9. If the auto alignment for the UUT has not been run within the last
hour, run it by pressing
System, Alignments, and Align All Now.
10.Ensure that the instrument passed all test before continuing.
Table 5-4 Signal Source Initialization
Parameter Setting
Frequency 50 MHz
Amplitude −25 dBm

Table of Contents

Other manuals for Agilent Technologies E4406A VSA Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4406A VSA Series and is the answer not in the manual?

Agilent Technologies E4406A VSA Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4406A VSA Series
CategoryTest Equipment
LanguageEnglish

Related product manuals