EasyManuals Logo

Agilent Technologies Truevolt Series User Manual

Agilent Technologies Truevolt Series
352 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #115 background imageLoading...
Page #115 background image
TEST:ALL?
Runs an instrument self-test and returns a pass/fail indication. The TEST:ALL? self-test is more comprehensive than
the *TST? self-test.
You must remove all input connections to the instrument before running the full self-test.
Parameter Typical Return
(none) +0 (pass) or +1 (one or more tests failed)
Run a self-test:
TEST:ALL?
l If one or more tests fail, the query returns +1 and stores an error in the error queue. For a complete listing of the
error messages related to self-test failures, see Self-test Error Messages.
l Following the test, the instrument returns to the instrument state that existed before the self-test.
Agilent Truevolt Series DMM Operating and Service Guide 115

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Truevolt Series and is the answer not in the manual?

Agilent Technologies Truevolt Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelTruevolt Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals