Calibration Process
The following general procedure is recommended to complete a full calibration.
1. Read Test Considerations.
2. Perform the verification tests to characterize the instrument (incoming data).
3. Unsecure the instrument for calibration (see Calibration Security).
4. Perform the adjustment procedures (see Adjustments).
5. Secure the instrument against calibration (see Calibration Security).
6. Note the security code and calibration count in the instrument’s maintenance records.
Agilent Truevolt Series DMM Operating and Service Guide 329