EasyManuals Logo

Agilent Technologies Truevolt Series User Manual

Agilent Technologies Truevolt Series
352 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #162 background imageLoading...
Page #162 background image
LXI Subsystem
This subsystem supports LAN eXtensions for Instrumentation (LXI) functionality.
Command Summary
LXI:IDENtify[:STATe]
LXI:MDNS:ENABle
LXI:MDNS:HNAMe[:RESolved]?
LXI:MDNS:SNAMe:DESired
LXI:MDNS:SNAMe[:RESolved]?
LXI:RESet
LXI:RESTart
LXI:IDENtify[:STATe] {ON|1|OFF|0}
LXI:IDENtify[:STATe]?
Shows or removes the LXI Web Identify indicator on the display.
Parameter Typical Return
{ON|1|OFF|0}, default OFF 0(OFF)or1(ON)
Turn on the LXI Web Identify indicator:
LXI:IDEN ON
l The indicator helps you identify the device associated with the LAN address.
l Press the [Local] key or send *RST to turn off the indicator.
LXI:MDNS:ENABle {ON|1|OFF|0}
LXI:MDNS:ENABle?
Disables or enables the multicast Domain Name System (mDNS), which provides the capabilities of a DNS server for
service discovery in a small network without a DNS server.
Parameter Typical Return
{ON|1|OFF|0}, default ON 0(OFF)or1(ON)
Turn mDNS ON:
LXI:MDSN:ENAB ON
l This setting is non-volatile; it is not changed by power cycling, a Factory Reset (*RST), or an Instrument Preset
(SYSTem:PRESet).
l This parameter is set to its default value when the instrument is shipped from the factory and after
SYSTem:SECurity:IMMediate.
162 Agilent Truevolt Series DMM Operating and Service Guide

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Truevolt Series and is the answer not in the manual?

Agilent Technologies Truevolt Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelTruevolt Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals