Frequency Accuracy Verification
Configuration: Frequency (CONFigure:FREQuency DEF, MIN)
1. Make sure you have read Test Considerations.
2. Select the Frequency function, default range, 3 Hz filter, and 1 second aperture.
3. Connect the Agilent 33500 Series function generator to the input terminals. For the 34461A, use the front-panel
input terminals and select the front input terminals with the Front/Rear switch.
4. Select each range in the order shown below. Provide the indicated input voltage and frequency. Compare
measurement results to the test limits shown below. (Be certain to allow for appropriate source settling.)
Values for 34461A DMM
Input Error from Nominal
Vrms Frequency Range Quick
Check
24 hours 90 days 1 year 2 years
0.1 Vrms 10 Hz 1 Vrms ± 3 mHz ± 3.5 mHz
0.01 Vrms 300 kHz 0.1 Vrms Q ± 60 Hz ± 180 Hz ± 300 Hz ± 450 Hz
Values for 34460A DMM
Input Error from Nominal
Vrms Frequency Range Quick
Check
24 hours 90 days 1 year 2 years
0.1 Vrms 10 Hz 1 Vrms ± 3 mHz ± 3.5 mHz
0.01 Vrms 300 kHz 0.1 Vrms Q ± 60 Hz ± 240 Hz ± 360 Hz ± 510 Hz
Q: Quick performance verification test point.
Agilent Truevolt Series DMM Operating and Service Guide 319

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Truevolt Series and is the answer not in the manual?

Agilent Technologies Truevolt Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelTruevolt Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals