CONFigure?
Returns a quoted string indicating the present function, range, and resolution. The short form of the function name
(CURR:AC, FREQ) is always returned.
Parameter Typical Return
(none) "VOLT +1.00000000E+01,+3.00000000E-06"
Return the present function, range, and resolution.
CONF?
CONFigure:CONTinuity
Sets all measurement parameters and trigger parameters to their default values for continuity measurements.
Parameter Typical Return
(none) +1.32130000E-02
Configure the instrument for continuity measurements.Then make a measurement using an external trigger with
positive slope (rising edge) and read the measurement:
CONF:CONT
TRIG:SOUR EXT;SLOP POS
READ?
l The range and resolution are fixed at 1 kΩ for continuity tests (a 2-wire resistance measurement).
l The instrument beeps (if the beeper is enabled) for each measurement less than or equal to the continuity
threshold (≤10 Ω), and the actual resistance measurement appears on the display.
l From 10 Ω to 1.2 kΩ, the instrument displays the actual resistance measurement with no beep. Above 1.2 kΩ,
the instrument displays "OPEN" with no beep.
l The FETCh?, READ?, and MEASure:CONTinuity? queries return the measured resistance, regardless of its value.
l Use READ? or INITiate to start the measurement.
Agilent Truevolt Series DMM Operating and Service Guide 139