EasyManuals Logo

Agilent Technologies Truevolt Series User Manual

Agilent Technologies Truevolt Series
352 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #139 background imageLoading...
Page #139 background image
CONFigure?
Returns a quoted string indicating the present function, range, and resolution. The short form of the function name
(CURR:AC, FREQ) is always returned.
Parameter Typical Return
(none) "VOLT +1.00000000E+01,+3.00000000E-06"
Return the present function, range, and resolution.
CONF?
CONFigure:CONTinuity
Sets all measurement parameters and trigger parameters to their default values for continuity measurements.
Parameter Typical Return
(none) +1.32130000E-02
Configure the instrument for continuity measurements.Then make a measurement using an external trigger with
positive slope (rising edge) and read the measurement:
CONF:CONT
TRIG:SOUR EXT;SLOP POS
READ?
l The range and resolution are fixed at 1 for continuity tests (a 2-wire resistance measurement).
l The instrument beeps (if the beeper is enabled) for each measurement less than or equal to the continuity
threshold (≤10 Ω), and the actual resistance measurement appears on the display.
l From 10 Ω to 1.2 kΩ, the instrument displays the actual resistance measurement with no beep. Above 1.2 kΩ,
the instrument displays "OPEN" with no beep.
l The FETCh?, READ?, and MEASure:CONTinuity? queries return the measured resistance, regardless of its value.
l Use READ? or INITiate to start the measurement.
Agilent Truevolt Series DMM Operating and Service Guide 139

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Truevolt Series and is the answer not in the manual?

Agilent Technologies Truevolt Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelTruevolt Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals