[SENSe:]{RESistance|FRESistance}:RANGe {<range>|MIN|MAX|DEF}
[SENSe:]{RESistance|FRESistance}:RANGe? [{MIN|MAX|DEF}]
Selects a fixed measurement range for all resistance measurements.
This parameter is common to 2-wire and 4-wire resistance measurements. Using the FRESistance version
of this command or query is identical to using the RESistance version.
Parameter Typical Return
{100 Ω|1 kΩ|10 kΩ|100 kΩ|1 MΩ|10 MΩ|100 MΩ}, default 1
kΩ
+1.00000000E+03
Configure 2-wire resistance measurements using the 10 kΩ range. Then make and read one measurement:
CONF:RES
RES:RANG 10E3
READ?
Typical Response: +6.27530000E+03
l Selecting a fixed range ([SENSe:]<function>:RANGe) disables autoranging.
l If the input signal is greater than can be measured on the specified manual range, the instrument displays the
word Overload on front panel and returns "9.9E37" from the remote interface.
See Also
[SENSe:]{RESistance|FRESistance}:RANGe:AUTO
Agilent Truevolt Series DMM Operating and Service Guide 207