STATus:QUEStionable[:EVENt]?
Returns the event register for the Questionable Data Register group. An event register is a read-only register that
latches events from the condition register. While an event bit is set, subsequent events corresponding to that bit are
ignored. The register bits are cleared when you read the register.
Parameter Typical Return
(none) +1024
Read the event register (bit 10 is set):
STAT:QUES?
l Once a bit is set, it remains set until cleared by reading the event register or by sending *CLS (clear status).