Values for 34460A DMM
Error from Nominal
Input Function
[1]
Range Quick
Check
24 hours 90 days 1 year 2 years
Open DC Current 100 µA Q ± 0.02 µA ± 0.025 µA
1 mA ± 0.060 µA
10 mA ± 2 µA
100 mA ± 4 µA ± 5 µA
1 A ± 60 µA ± 100 µA
3 A ± 600 µA
Short DC Volts 100 mV Q ± 6 µV ± 6.5 µV
1 V ± 9 µV ± 10 µV
10 V ± 40 µV ± 50 µV
100 V ± 600 µV
1000 V ± 6 mV ± 10 mV
Short 4-wire Ohms 100 ± 6 ± 7 m
1 kΩ ± 8 mΩ ± 10
10 kΩ Q ± 50 ± 100 mΩ
100 kΩ ± 500 ± 1 Ω
1 MΩ ± 10 Ω
10 MΩ ± 100 Ω
[1]
Select 10 NPLC.
Q: Quick performance verification test point.