Values for 34460A DMM
Input Error from Nominal
Vrms Frequency Range Filter Quick
Check
24 hours 90 days 1 year 2 years
100 mV 1 kHz 100 mV 200 Hz ± 90 µV ± 110 µV ± 120 µV ± 130 µV
50 kHz ± 170 µV ± 190 µV ± 200 µV ± 210 µV
300 kHz Q ± 4.5 mV
1 V 1 kHz 1 V 200 Hz ± 900 µV ± 1.1 mV ± 1.2 mV ± 1.3 mV
50 kHz ± 1.7 mV ± 1.9 mV ± 2.0 mV ± 2.1 mV
300 kHz ± 45 mV
0.03 V 1 kHz 10 V 200 Hz ± 2 mV ± 3 mV
1 V 1 kHz 200 Hz ± 2.7 mV ± 3.8 mV ± 3.9 mV ± 4.0 mV
10 V 10 Hz 3 Hz ± 9 mV ± 11 mV ± 12 mV ± 13 mV
100 Hz 20 Hz ± 9 mV ± 11 mV ± 12 mV ± 13 mV
20 kHz 200 Hz ± 9 mV ± 11 mV ± 12 mV ± 13 mV
50 kHz 200 Hz Q ± 17 mV ± 19 mV ± 20 mV ± 21 mV
100 kHz ± 66 mV ± 71 mV
300 kHz ± 450 mV
100 V 1 kHz 100 V 200 Hz Q ± 90 mV ± 110 mV ± 120 mV ± 130 mV
50 kHz ± 170 mV ± 190 mV ± 200 mV ± 210 mV
70 V 300 kHz ± 3.3 V
750 V 1 kHz 750 V 200 Hz ± 675 mV ± 825 mV ± 900 mV ± 975 mV
210 V 50 kHz ± 573 mV ± 669 mV ± 690 mV ± 711 mV
70 V 300 kHz ± 6.6 V
Agilent Truevolt Series DMM Operating and Service Guide 321