Fault removal
118 Edition 01/2016 multi EA 4000
9.6.5 Measurements in pyrolysis mode (EC)
Results of elementary carbon too low (after pyrolysis)
Cause
Solution
Waiting time to display the oxygen
with argon in the combustion tube
chosen too short
Waiting time after changing from oxygen to
argon supply should be at least 60 s.
Results of elementary carbon too high (after pyrolysis)
Cause Solution
Pyrolysis time selected too short
The pyrolysis time should be at least 3 min. to
ensure that all organic compounds have been
dissolved and purged completely.
Pyrolysis temperature chosen too
high
At temperatures above 1000 °C soot is in-
creasingly formed from organic compounds
(plastics) through "coking", which is also de-
tected in stage 2 as elementary carbon. The
optimum furnace temperature for pyrolysis
measurements is 850 °C.