Operation
70 Edition 01/2016 multi EA 4000
Repeat the measurement of the above-mentioned standard 3 times and determine
the average value.
Calculate the splitting factor F:
… expected absolute Cl content in µg
… average absolute Cl content from the 3 standard measurements
Enter the calculated factor in the method on tab CALIBRATION in the input field DAILY
FACTOR
and accept the values with [ACCEPT] for the subsequent measurements.
IMPORTANT
To calculate a new splitting factor the
DAILY FACTOR
must be set to a value of 1 in the method
on the tab
CALIBRATION .
7.4.7 Cauterizing the quartz boat
Quartz boats have to be cauterized
before first use
after cleaning out heavy contamination.
To this end perform the measuring routine with empty quartz boats.