Transport and storage
multi EA 4000 Version 01.16 119
10 Transport and storage
10.1 Transport
10.1.1 Preparing the analyzer for transport
CAUTION! RISK OF BURNING AT THE HOT FURNACE!
There is a risk of burns at the hot combustion furnace. Only remove the combustion tube
when the device is cold or allow the device to cool down sufficiently.
CAUTION! RISK OF INJURY FROM GLASS BREAKAGE!
When removing the glass components there is a risk of injury from glass breakage! Remove
all glass components carefully from the analyzer!
BLE TRANSPORT PACKAGING!
Unsuitable packaging material and resid
ue of measuring solution and chemicals can dam-
age individual components of the analyzer!
Only transport the analyzer in its original packaging! Ensure that the analyzer is fully drained
and all transport locks have been fitted!
Prepare the EA 4000 and its modules for transport as follows:
Basic device multi EA 4000
1. Switch off the analyzer from the main switch and allow the equipment to cool down.
2. Disconnect the gas supply and unplug the mains plug from the mains outlet.
3. Undo all connections on the back of the analyzer.
4. Open the doors of the analyzer. Empty the drying tube. Reinsert the drying tube.
5. Unscrew the connections at the halogen trap and press the halogen trap out of the
clamps.
6. Pull the combustion tube out of the furnace.
7. Carefully package the accessories, in particular protect glass components against
breakage.
C/S module
1. Switch the C/S module off from the main switch.
2. Disconnect the power cable from the mains socket.
3. Undo all connections on the back of the module.
Cl module
1. Switch the Cl module off from the main switch.
Remove the measuring cell from the device and drain it (see section "