Transport and storage
multi EA 4000 Version 01.16 121
For the setup at the new location observe the notes in section "Site requirements" p.
49.
10.2 Storage
CAUTION
Environmental influences and condensate formation can destroy individua
l components of
the analyzer!
The analyzer must only be stored in
air-conditioned rooms. The atmosphere must be low in
dust and free from aggressive vapors.
If the analyzer and add-on devices are not positioned immediately after delivery or are not
required for a prolonged period of time, they should best be stored in their original packag-
ing. A suitable desiccant should be added to the equipment to prevent damage from mois-
ture.
The following requirements are placed on the climatic conditions in the storage room of the
analyzer:
Temperature range: +5 °C to +55 °C
max. humidity: 10 % to 30 %
Air pressure: 0.7 bar to 1.06 bar