Fault removal
116 Edition 01/2016 multi EA 4000
9.6.2 Cl detection
IMPORTANT
To prevent errors always note:
Connect measuring cell with stirrer rod (filled with electrolyte) and electrode before
switching on the basic device and Cl module.
Start the multiWin program approx. 3 min after switching on the analyzer.
Recovery too low (splitting factor > 1.4)
Cause
Solution
sulfuric acid too diluted check and replace sulfuric acid
components or measuring gas
hoses sooted up
Check combustion tube for contamination.
Clean the combustion tube by burning off the
deposits with a Bunsen burner or in the muffle
furnace at 1000 °C. Replace heavily contami-
nated tubes. Replace the quartz wool in the
combustion tube
Check sulfuric acid container for soot. If parti-
cles have entered as far as the sulfuric acid
container, clean the container and connection
at the furnace outlet and refill the container.
Recovery too high (splitting factor < 0.6)
Cause Solution
Sample deposits in the combustion
tube (e. g. salts)
Replace or clean combustion tube, e.g. by
"soaking" the combustion tube overnight in
10% HNO
3
, then rinse thoroughly with distilled
water and dry.
Shutdown threshold below 1500
Cause Solution
Combination electrode faulty or too
old
Use new electrode for chlorine detection.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Analytik Jena multi EA 4000 and is the answer not in the manual?

Analytik Jena multi EA 4000 Specifications

General IconGeneral
BrandAnalytik Jena
Modelmulti EA 4000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals