Safety instructions
20 Edition 01/2016 multi EA 4000
3.7.6 Safety instructions - service and repair
The analyzer is usually serviced by the customer service department of Analytik Jena AG or
its authorized and trained specialist personnel.
Independent servicing can maladjust or damaged the analyzer. Therefore, the operator may
generally only carry out the tasks listed in chapter 'Maintenance and care' p. 84.
The following has to be observed:
The exterior of the analyzer may only be cleaned with a damp, not dripping, cloth af-
ter the analyzer has been switched off.
Any service and repair work at the analyzer may usually only be carried out in the
switched-off condition (unless stated otherwise).
Service tasks and the replacement of system components (removal of the combus-
tion tube, service of the dust trap) must only be carried out after a sufficiently long
cooling down phase.
Prior to servicing or repair the energy and gas supplies must be disconnected and
the analyzer must be v vented!
Only use original accessories and original replacement parts from Analytik Jena AG.
The notes in the chapter "Service and care" must be observed.
All protective equipment must be reinstalled correctly immediately after completion of
the service and repair work and be checked for operation!
3.8 Behavior during emergencies
The analyzer must be switched off from the main switch (in the multi EA on the front panel, in
the modules on the equipment backplate) and the mains plug must be disconnected from the
mains outlet in case of dangerous situations or accidents.
Because a rapid response can save lives during an emergency, the following has to be en-
sured:
The operating staff must be familiar with the location of safety equipment, accident
and danger alarms as well as first aid and rescue equipment as well as their han-
dling.
The operator is responsible for the respective training of the operating staff.
All equipment for first aid (first-aid kit, eyewash bottles, stretcher, etc.) as well as
equipment for firefighting (fire extinguishers) must be within reach and easy to ac-
cess. All equipment has to be in a sound condition and should be checked regularly.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Analytik Jena multi EA 4000 and is the answer not in the manual?

Analytik Jena multi EA 4000 Specifications

General IconGeneral
BrandAnalytik Jena
Modelmulti EA 4000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals