System design
38 Edition 01/2016 multi EA 4000
4.6 Manual TIC solids module
The manual TIC solids module enables the detection of TIC in solids in combination with the
multi EA 4000 C or multi EA 4000 C/S. The acidification and purging of the TIC takes place
at raised temperatures. The measuring gas is moved together with the carrier gas to the
detector of the C/S module.
The manual TIC solids module is only designed for the manual sample supply. Automation
using a sampler is not possible.
1 manual TIC solids module
2
C/S detector
basic device multi EA 4000
Fig. 24 Manual TIC solids module
4.6.1 Technical data
Designation/type Manual TIC solids module
Dimensions 300 x 500 x 550 mm
Mass approx. 10 kg
Procedural data
Measuring method TIC
Detection NDIR (coupled with VITA method)
Sample volume Up to 3000 mg
Gas supply Oxygen min. 2.5
2 – 4 bar
Connection 230 V AC (+ 10%, -15%), 50/60 Hz
Protection T 16 A H
Typical average power con-
sumption
3700 VA