EasyManua.ls Logo

Analytik Jena multi EA 4000 - Analytical Problems; C;S Detection

Default Icon
148 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Fault removal
multi EA 4000 Version 01.16 115
9.6 Analytical problems
9.6.1 C/S detection
Recovery generally too low
Cause
Solution
Drying agent too moist
Filter and measuring gas hoses
sooted up
Check and replace the desiccant
Check and replace the filter
Check dust trap, if heavily contaminated (par-
ticle deposits), clean dust trap and replace
measuring gas hose up to the filter
Too low results in TS
Cause
Solution
Halogen trap still in measuring gas
flow
Replace halogen trap with an empty glass
tube
Digestion temperature too low
(especially for sulfate bonding
form)
Increase furnace temperature in the method,
if S measuring graph look very irregular or
has a long trailing and does not reduce to the
base line.
If necessary, add additives (e. g. iron phos-
phate, tungsten oxide, vanadium pentoxide,
tin…).
Contamination of the hoses carry-
ing the measuring gas up to the
optics
Remove hoses up to the NDIR optics, clean,
dry and re-install them.
Note:
We recommend to have the cleaning of the
hoses carried out by Service.
Excess results during TC
Cause Solution
Combustion tube has a crack or is
broken
Ambient air can enter through the cracks into
the system and may cause significant excess
results in TC. Replace combustion tube
Base line drift
Cause
Solution
Detector not stable
The detector must stabilize during the warm-
up phase of the device. Therefore, it may
happen at the start of measuring that the base
line still drifts a little (temperature depend-
ence). The peak area changed due to the drift
will be corrected automatically, thus the
measuring results are never affected.
If the drift does not stop, inform Service.

Table of Contents

Related product manuals