Page #38 background image
3-19
MS9710C Device Message List (12/13)
Item
Command
Data
request
Response
TMK λ
λ = xxxx.xxxx
DMK λ
λ = xxxx.xxxx
MKA λ
MKB λ
λ = xxxx.xxxx
MKC l
MKD l
l : Valueshown
on the right
EMK
ZMK WL,
λc, λs
λc = xxxx.xxx
λs = xxxx.xxx
ZMK SPN
ZMK ZOOM, s
s=IN
= OUT
ZMK ERS
SSI
SRT
SST
AUT
OPT s
s = ON, OFF
TMK?
DMK?
MKA?
MKB?
MKC?
MKD?
ZMK?
WL
ZMK?
ZOOM
AUT?
OPT?
λ, l
λ =
xxxx.xxxx
λ =
xx.xx (dBm , dB)
=
Four signifi cant
digits (*W, %)
∆λ, l
∆λ = xxxx.xxxx
l = xx.xx (dB)
= xxx.xxx
λ
λ
λ = xxxx.xxxx
l
l
l=
xx.xxx (dBm, dB)
=
Seven significant
digits (*W, %)
WL, λc, λs
λc = xxxx.xxx
λs = xxxx.xxx
ZOOM, s
s=IN
= OUT
n
n=
0: Measurement
end
= 1:
During
measurement
s
s = ON, OFF
Device message
Remarks
Marker
Zone
Marker
Sweep
λ : Wavelength (nm) or
(THz)
l : Level
dBm, dB, *W, %
∆λ : Difference in
wavelength (nm)
or (THz)
l : Difference in level
Log dB, no linear unit
λ Wavelength (nm) or
(THz)
l : Level
dBm, dB, *W, %
λc, λs : Unit (nm)
λc : Zone center
λs : Zone span
Option
9.87
9.27
9.52
9.53
9.54
9.55
9.32
9.96
9.76
9.75
9.59
9.9
9.60
Auto Measurement
Light Output
Trace Marker
Marker
Wl Marker A
B
Lvl Marker C
D
Marker Off
Zone Marker
ZoneSpan
Zoom
In/Out
Zone Marker
Erase
Single
Repeat
Stop
3.3 Device Message List

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Anritsu MS9710C and is the answer not in the manual?

Anritsu MS9710C Specifications

General IconGeneral
BrandAnritsu
ModelMS9710C
CategoryMeasuring Instruments
LanguageEnglish