5-7
5.5 A05 MOD DSP1
5.5.1 Preparation
Make all the variable resistors at the center of the variable range. (R112, R115, R124,
R128, R132)
Push RX measure key in PHS mode.
Set RX parameter
Burst=off
Pattern=0000
Set in Bit error rate screen
RF Output=on
Modulation=on
5.5.2 “I” Level adjustment
Adjust R115 to become 4.0Vp-p ±0.2 V on oscilloscope.
5.5.3 Offset adjustment of “I,Q”
Measure RF output with a spectrum analyzer.
Adjusts R112 and R118 for the level to become minimum (below 50 dB than
carrier) at the carrier frequency+24 kHz.
5.5.4 Balance and phase of “I/Q” adjustment
Measure RF output with a spectrum analyzer.
Adjusts R132 and R124 for the level at the carrier frequency24 kHz to become
minimum (below 50 dB than the level at the carrier frequency+24 kHz).
5.5 A05 MOD DSP1

Table of Contents

Related product manuals