7-9
7.5 Outline of Displayed Screen
7.5.5 Sensor Data Input (Power Sensor and Calibration Factor Input)
Screen
When a measurement is to be performed using a power meter (including prior calibra-
tion), the “Power Sensor and Calibration Factor Input Screen” shown in Fig. 7-5 is
displayed in this program to enable the selection of a sensor and to set the calibration
factor.
The following section describes the name and gives a brief summary of the function of
each control effective on this screen.
[1] [2] [3] [4] [5] [6] [7][8][9] [10] [11]
Fig. 7-5 Power Sensor and Calibration Factor Input Screen
[1] “Data_Select” bar Used for the selection and new registration of the power sensor equipment name. See
paragraph 7.6.3 “ Registering Power Sensor and Setting Calibration Factor” for more
information.
[2] Amorphous Power Sensor (Amorphous sensor data display field)
Displays the data on the currently set sensor when a measurement is performed using
the amorphous power sensor. Editing is performed in item [11] “Data Input”.
[3] Serial number display field Displays the equipment name of the sensor for each sensor type. Changes are made in
item [1] “Data_Select”
[4] Reference Cal Factor (Reference calibration factor display field)
Displays the data on the reference calibration factor currently set for each sensor type.
Changes are made via direct key input.