5-9
5.6.3 AF Output level adjustment
(1) Setting of measurement system
MT8802A
Analog measure-AF measure
AF Osc.1 Freq.=1 kHz, Level=2V
AF Osc.2=off
AF Output Impedance=600
HP8903 Audio Analyzer
LPF=30 kHz
Connect the MT8802A AF Output to the HP8903 AF input (100 k ohms) termi-
nal.
(2) Adjust the R100 (OSC variable resistor) so that the HP8903 measurement result
(AF level) is within 1.988 Vrms ±0.5%.
5.6.4 AF Osc.2 level adjustment
(1) Setting of measurement system
MT8802A
Analog measure-AF measure
AF Osc.1=off
AF Osc.2 Freq.=1 kHz, Level=2V
AF Output Impedance=600
HP8903 Audio Analyzer
LPF=30 kHz
Connect the MT8802A AF Output to the HP8903 AF input (100 k ohms) termi-
nal.
(2) Adjust the R61 (DA2 variable resistor) so that the HP8903 measurement result
(AF level) is within 1.988 Vrms ±0.5%.