5-13
5.8 MT8802A Software Operation Instructions
5.8.1 Preface
5.8.1.1 Outline
This procedure instructs the operation (confidential) required for delivery inspection in
the MT8802A Software Inspection Department, and for maintenance in the Service
Department. When a sales personnel or user upgrades the version, transfer the
MT8802A Software Version Upgrade Handbook of the separate volume.
When delivery inspection errors are detected in this procedure, they must be improved
each time they are detected.
5.8.1.2 Operation items
This procedure explains the operation items below.
Setting DIP switches on front panel
Registered master file
Software installation
Writing basic calibration data
Setting Spectrum option
Serial number management
Clearing all of SRAM
Setting parameters to the state of shipment from the factory
5.8.2 Setting DIP Switches on Front Panel
6 5 4 3 2 1
O N
DIP switches
Setting at factory shipment
OFF
OFF
OFF
OFF
OFF
OFF
Meaning for ON
Software debug mode
Basic calibration data is not set on the
measurement result.
Not used.
Not used.
Can control the power-on/off with the
switch on the rear panel.
Not used.
bit
1
2
3
4
5
6
5.8.2.1 Software debug mode
Displays messages at installation.
Displays detail message at error occurrence.
Can set/clear the serial number.
Can install a new software option.
Can format the ATA card forcibly.
5.8 MT8802A Software Operation Instructions