EasyManua.ls Logo

Anritsu MT8802A - Page 60

Anritsu MT8802A
152 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Section 6 Correction Software
6-12
Input the serial number by following the instructions below.
(1) Select a measurement item in section [2] on the Initial Menu Screen shown in
Fig. 6-4. When Receiver is selected in this section and Channel Selector is
selected in section [19], the Connect box of section [11] is displayed, enabling
setting in the range from one to four. On the other hand, when Transmitter is
selected, continuous measurement of multiple DUTs is not enabled. Therefore,
section [11] is not displayed.
(2) Select Settings in section [1] and Serial Number in section [17] on the same
screen. This causes the Serial number Input Screen, shown in Fig. 6-6, to be
displayed. When continuous measurement of multiple units is selected in (1),
sequential numbers starting from 1 are allocated in Address Number in section
[1].
(3) Input a serial number that corresponds to each address number in the Serial
Number Input box in section [3]. Input of up to 8-digit alphanumeric characters
is only allowed.
(4) Click the OK button in section [4] when you have completed inputting all the
serial numbers. This causes the screen to return to the Initial Menu Screen
shown in Fig. 6-2. When this button is selected before you complete inputting all
the serial numbers, a message as shown in Fig. 6-7 appears. To exit from this
screen in the process of inputting procedures, select the Cancel button. When a
serial number is found to duplicate, a message as shown in Fig. 6-8 appears. In
this case, check the input serial number and correct either one of the duplicated
numbers.
Fig. 6-7 Message Requesting Confirmation of Serial Number Input
Fig. 6-8 Message Requesting Confirmation of Serial Number Duplicated

Table of Contents

Related product manuals