Chapter 4: Maintenance 113
12 Click OK when monthly cleaning is complete.
NOTICE If you have been running BD FACSRinse solution as part of a biweekly
cleaning procedure, you can repeat the monthly cleaning using BD FACSRinse
solution followed by DI water.
13 Disconnect the tank, fill it with sheath solution, and reconnect it.
14 Reconnect the BD LSR II sheath filter.
NOTICE If you are performing a sheath fluid exchange, install a new filter.
15 Check for bubbles and prime, if necessary.
BD FACSCanto and BD FACSCanto II Monthly Cleaning
1 Ensure the sample coupler is properly installed and the HTS is powered on.
2 Empty the waste tank, if necessary.
3 Ensure the BD FACSClean and BD FACS shutdown solution cubitainers
are full and connected.
4 Choose Cytometer > Long Clean.
5 Click OK when the Long Clean is complete.