164 BD High Throughput Sampler User’s Guide
For details on removing the probe, see Replacing the Probe on page 129.
Figure B-3 Removing the sample probe and tubing
2 Pipette a 10% bleach solution into the injection port/wash station.
Using a disposable pipette, fill the injection port/wash station completely,
being careful not to overfill.
3 Connect the line declogger tool to the waste connector on the back of the unit.
Figure B-4 Attaching the declogger tool
All cytometer surfaces that come in contact with biological specimens can
transmit potentially fatal disease. Treat the sample probe and tubing as
biohazardous waste and dispose of them according to local regulations.
sample probe
sample probe tubing