Appendix B: Depot Repair Procedures 165
4 Slowly pull back the plunger of the declogger tool until all of the bleach
solution is drawn into the barrel.
5 Disconnect the declogger tool and discard the bleach solution according to
local regulations.
6 Repeat step 2.
7 Reattach the declogger tool, and slowly remove about half of the bleach
solution; allow the remaining bleach to sit for 30 minutes.
Keep the declogger tool attached to the waste port during the 30 minutes.
8 Use the declogger tool to remove the rest of the bleach solution.
9 Disconnect the declogger tool and discard the bleach solution according to
local regulations.
Rinsing the Fluidics
1 Pipette DI water into the injection port/wash station.
Using a disposable pipette, fill the injection port/wash station completely,
being careful not to overfill.
2 Connect the line declogger tool to the waste connector on the back of the
unit.
3 Slowly pull back the plunger of the declogger tool until all of the DI water
is drawn into the barrel.
4 Repeat step 1 and step 3.
5 Disconnect the declogger tool and discard the DI water.
6 Carefully dry the declogger tool and return it to the HTS accessory kit.
7 Remove and discard the secondary pump valve tubing (A).