24 BD High Throughput Sampler User’s Guide
6 moves the probe up, over, and down
into the injection port
moves the probe up, over, and down
into the injection port
7 injects the sample and boosts it to
the cytometer flow cell at high
speed
a non-selectable, fixed boost
command transports sample to the
flow cell to prime the sample flow
path
injects the sample and boosts it to
the cytometer flow cell at high
speed
a non-selectable, fixed boost
command transports sample to the
flow cell to prime the sample flow
path
8 slows down the sample to the
required acquisition speed and
acquires the required sample volume
delivers the sample to the flow cell
at the required rate and volume
using the secondary (high-
throughput) pump
simultaneously, raises the probe to
the cleaning position and the
primary pump begins processing
the next well in parallel to
acquisition
9 pushes any remaining sample and the
separator bubble through the flow
cell, and a new cycle begins
once the secondary pump delivers the
sample, pushes any remaining sample
and the separator bubble through the
flow cell. The next well’s sample is
injected and a new cycle begins
10 after the last well in a sequence, raises
the probe to the cleaning position and
performs a final flush and refill
after the last well in a sequence, raises
the probe to the cleaning position and
performs a final flush and refill
11 lowers the probe into the injection
port/wash station
lowers the probe into the injection
port/wash station
Sequence Standard Mode High-Throughput Mode