12 - MC6-T Safety
Operating
Temperature
0 … 45 °C (32 … 113 °F)
Storage Temperature -20 ... 60 °C (-4 ... 140 °F)
Operating Humidity 0 to 90 % R.H. non condensing
Storage Humidity 10 to 60 % R.H. non condensing
Max. safe operating
altitude
5000 m (approx. 16404 ft)
Pollution Degree 2 (only non-conductive POLLUTION occurs except
that occasionally a temporary conductivity caused
by condensation is expected)
Indoor use only
Note: Beamex provided temperature specifications are valid for
ambient temperatures between 13..33 °C. Outside of this scope,
temperature coefficients must be used. For more information
about temperature coefficients see User Guide for Beamex CMX
Calibration Software.
Note: MC6-T150's heating/cooling modules undergo normal
aging process during usage, after which minimum reachable
calibration temperature may be higher.
Note: Using the calibrator in a highly pollutive environment,
such as coal mine and steel factory is not allowed, unless
the calibrator is stored and used in a non-pollutive area, for
example a dedicated calibration room that has suitable air filters.
Contamination of conductive dust can cause electrical hazard.
Note: If the calibrator has been stored in different environment,
it should be allowed to stabilize to new environment before use.
For all measurement ports except IN and OUT a cable not
longer than 3 m must be used.
Sometimes it is necessary to use a portable radio transceiver
while working with the calibrator. To prevent calibration errors
caused by the radio frequency interference, keep the radio
far (at least 1 meter) from the calibrator and the circuit under
calibration while sending.