EasyManua.ls Logo

BEAMEX MC6-T - Overtemperature Protection Test Mode

BEAMEX MC6-T
172 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
142 - Maintenance
Overtemperature Protection Test Mode
MC6-T is equipped with an additional safety feature called
Overtemperature Protection Test Mode. It gives a possibility to make
sure that the power is disconnected from the Temperature Block in case it
heats up over the maximum setpoint temperature.
To enter the Overtemperature Protection Test mode select Settings
from the Home view. Tap Maintenance button, select Change module
data option and provide the required PIN code: 5656. After accepting the
changes the device is in test mode.
Figure 139: Overtemperature Protection Test Mode window
To test the feature, enable the temperature generation to maximum
setpoint specific for your model. If the protective mechanism works
properly, you will be prompted with a message shown in Figure 140:
Overtemperature Protection Test Mode prompt message. The power will
be disconnected from the Temperature Block and the device will start
cooling. To return to the normal operating mode, switch the Mains switch
off and then back on, to clear the error condition.
CAUTION: If the maximum setpoint temperature stabilizes
without an error message being shown, contact Beamex. Safe
use of MC6-T is no longer ensured.