Warnings Concerning Temperature - 15
Warning: MC6-T's handle must be in down position during
calibration to prevent it from getting too hot.
After using the Temperature Block
Warning: If the Temperature Block has been heated up to
temperatures above 50 °C, it must be cooled down below this
temperature before switching off the device.
Warning: If the Temperature Block has been cooled down to
temperatures below dew point, it must be heated up until all
water has evaporated from the Temperature Block and insert.
Temperature Block must then be cooled down below 50 °C
before switching off the device.
Warning: Always remove the insert from the Temperature Block
when not using the calibrator. Use the insert removal tool for
removing the insert.
Warning: Always switch off the Mains switch when not using
the device.
Warning: Never set the device inside a carry case if the
Temperature Block temperature is above 50 °C or below dew
point.