56 - Active Terminals and Connections
Figure 42: Frequency generation terminals. Range 0.0005 … 50000 Hz
See also Frequency Measurement and Pulse Generation.
Pulse Generation
Before generating pulses, the following settings should be checked:
Frequency. To set the frequency, tap on the button with the "Hz" value.
Amplitude. Defined from the button with the "V" value.
Waveform and Duty Cycle. Set together from the rightmost button.
Duty Cycle is the ratio of the output high time to the total cycle time. Due
to technical reasons, the entered Duty Cycle is not always realized with
relatively high frequencies. When the realized Duty Cycle differs from the
desired Duty Cycle, the realized Duty Cycle is shown with an asterisk (*) in
front of it, e.g.:
Figure 43: Pulse generation terminals. Range 0 … 9 999 999 pulses
See also Pulse Counting and Frequency Generation.