Patient monitor Users manual
17-2
17.1.2 Calculation Unit
Drug A, drug B, drug C, drug D, drug E are not the real name of drugs, but
only code name of drugs. The five kinds of drugs units are fixed. You can
select the appropriate units according to the physicians' comments. The rules
of unit displaying are as follows:
Drug A, drug B and drug C are fixed to g series units: g, mg and mcg.
Drug D is fixed to unit series units: unit, k unit and m unit.
Drug E is fixed to mEq unit.
When define some kind of drug by yourself, you should select drug A,
drug B, drug C, drug D, drug E according to the unit series.
Caution: Drip speed and volume per drip are invalid for neonatal.
17.1.3 Titration Table
After finishing the calculation of drug dose calculation, select
Titration in the interface of drug dose calculation to enter the interface of
titration table.
You can rejigger the following options in the titration table:
ReferenceYou can select fromDose,INF RateandDrip
Rate.
Dose Type: Select the dose unit according to your need.
Step: You can select from 110.
The data in the titration table will have some change after finishing the
above options.
You can also select or and turn page up and down to browse more
data. Select buttonRecord, it will output the titration table data in the
current screen. DOSAGE means dose while SPEED means transfusion speed.