EasyManuals Logo

Biolight Q7 User Manual

Default Icon
210 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #176 background imageLoading...
Page #176 background image
Patient monitor Users manual
10
Channels Default : II lead
Measurement range -2.0 mV to +2.0 mV
Accuracy
-0.8 mV to +0.8 mV: ±0.02 mV or ±10%, whichever is
greater; Over ±0.8mV: unspecified
Resolution 0.01mV
Alarm range
-2.0 mV to +2.0 mV, high/low limit can be adjusted
continuously.
Arrhythmia analysis
Kinds
ASYSTOLE, VENT FIB, VPB, COUPLET, MULTI PVCS,
BIGEMINY, TRIGEMINY, R ONT, MISSED BEATS, ST
HIGHER, ST LOWER,TACHY, BRADY, PNC, PNP,
NOISE, VTACHY, PVCS HIGHER
ECG/ST/arrhythmia supplemental information as required by AAMI EC11/13
Electrosurgery
protection
electrocautery uint
protection
Cut mode: 300W
Coag mode: 100W
HR change: ≤10%
Resuming time: ≤10s
Complies with ANSI/AAMI EC 13:2002, 4.2.9.14
Input circuit current < 0.1uA
Tall T-Wave rejection
capability
Minimum recommended 1.2 mV T-Wave amplitude
Complies with ANSI/AAMI EC 13:2002, 4.1.2.1 c)
Heart rate averaging
≤ 50 bpm, once every 4 beats;
> 50 bpm, once every 12 beats.
HR meter accuracy
and response to
irregular rhythm
Ventricular bigeminy: 80 bpm
Slow alternating ventricular bigeminy: 60 bpm
Rapid alternating ventricular bigeminy: 120 bpm
Bidirectional systoles: 90 bpm
Response time of HR
meter to change in HR
HR change from 80 bpm to 120 bpm: less than 10s
HR change from 80 bpm to 40 bpm: less than 10s
Complies with ANSI/AAMI EC13-2002: 4.1.2.1 f

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Biolight Q7 and is the answer not in the manual?

Biolight Q7 Specifications

General IconGeneral
BrandBiolight
ModelQ7
CategoryMedical Equipment
LanguageEnglish

Related product manuals