Patient monitor User’s manual
10
Channels Default : II lead
Measurement range -2.0 mV to +2.0 mV
Accuracy
-0.8 mV to +0.8 mV: ±0.02 mV or ±10%, whichever is
greater; Over ±0.8mV: unspecified
Resolution 0.01mV
Alarm range
-2.0 mV to +2.0 mV, high/low limit can be adjusted
continuously.
Arrhythmia analysis
Kinds
ASYSTOLE, VENT FIB, VPB, COUPLET, MULTI PVCS,
BIGEMINY, TRIGEMINY, R ONT, MISSED BEATS, ST
HIGHER, ST LOWER,TACHY, BRADY, PNC, PNP,
NOISE, VTACHY, PVCS HIGHER
ECG/ST/arrhythmia supplemental information as required by AAMI EC11/13
Electrosurgery
protection
(electrocautery uint
protection)
Cut mode: 300W
Coag mode: 100W
HR change: ≤10%
Resuming time: ≤10s
Complies with ANSI/AAMI EC 13:2002, 4.2.9.14
Input circuit current < 0.1uA
Tall T-Wave rejection
capability
Minimum recommended 1.2 mV T-Wave amplitude
Complies with ANSI/AAMI EC 13:2002, 4.1.2.1 c)
Heart rate averaging
≤ 50 bpm, once every 4 beats;
> 50 bpm, once every 12 beats.
HR meter accuracy
and response to
irregular rhythm
Ventricular bigeminy: 80 bpm
Slow alternating ventricular bigeminy: 60 bpm
Rapid alternating ventricular bigeminy: 120 bpm
Bidirectional systoles: 90 bpm
Response time of HR
meter to change in HR
HR change from 80 bpm to 120 bpm: less than 10s
HR change from 80 bpm to 40 bpm: less than 10s
Complies with ANSI/AAMI EC13-2002: 4.1.2.1 f