Patient monitor Users manual
15
continuously.
Recovery time after
defibrillation
<5s
Nellcor SpO
2
SpO
2
Measurement range 0% to 100%
Resolution 1%
Accuracy
70% to 100%: ±2%adult/pediatric
70% to 100%: ±3%neonate
0% to 69%, unspecified
Alarm range
0% to 100%, high/low limit can be adjusted continuously.
Average time 8s, 16s
PR
Measurement range 20 bpm to 300 bpm
Accuracy
20 bpm to 250 bpm: ±3 bpm
251 bpm to 300 bpm: unspecified
Resolution 1 bpm
Alarm range
0bpm300bpm, high/low limit can be adjusted continuously.
Recovery time after
defibrillation
<5s
Masimo SpO
2
SpO
2
Measurement range 0% to 100%
Resolution 1%
Accuracy
70% to 100%:±2%adult/pediatric, non-motion conditions
70% to 100%:±3%neonate, non-motion conditions
70% to 100%:±3%motion conditions
0% to 69%,unspecified
Average time 2-4s, 4-6s, 8s, 10s, 12s, 14s, 16s
PR
Measurement range 25 bpm to 240 bpm