EasyManuals Logo

Biolight Q7 User Manual

Default Icon
210 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #71 background imageLoading...
Page #71 background image
Patient Monitor Users manual
7-7
7.3.3 Connecting ECG Cable
Plug the ECG cable into the ECG connector. An ECG waveform and numeric
appears on the monitor display.
7.3.4 Selecting Leads
Depending on the patient to be monitored, you shall select the proper leads
as required.
1. Select ECG parameter area, enter ECG parameter setup menu, selectLead
Type.
2. You can select3 lead,5 leador12 lead.
7.3.5 Checking Paced Status
It is important to set the paced status correctly when you start monitoring
ECG.
If thePacedstatus has been set toYes, the icon will be shown
on the screen. If the system detects paced signal, the symbol
will be
marked on the ECG wave.
To change the paced status, you can selectPatient Manage.smartkey
Pat. Information, and set thePacedstatus in the popped up menu.
Warning:
For paced patients, you must setPacedtoYes. If it is incorrectly
set toNo, the monitor could mistake pace pulses for regular QRS
complexes and fail to alarm during asystole.
Some pace pulses can be difficult to reject. When this happens, the
pulses are counted as a QRS complex, and could result in an incorrect
HR and failure to detect cardiac arrest or some arrhythmias. Keep
pacemaker patients under close observation.
For non-paced patients, you must setPacedtoNo, otherwise, the
system can not detect the arrhythmia related to Ventricular
Premature(including PVCs count), and will not conduct ST analysis.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Biolight Q7 and is the answer not in the manual?

Biolight Q7 Specifications

General IconGeneral
BrandBiolight
ModelQ7
CategoryMedical Equipment
LanguageEnglish

Related product manuals