Patient Monitor Users manual
7-6
Warning:
To avoid burning, when the electrotome operation is performed, the
electrodes should be placed near the middle between ESU grounding
pad and electrotome and the electrotome should be applied as far as
possible from all other electrodes, a distance of at least 15 cm/6 in is
recommended.
When using the ESU device, avoid placing the electrodes near the ESU
grounding pad, otherwise, grate deal interference will influence the
ECG signals. The monitor should be placed far from the operating
table. Power wires and the ECG cables should be partitioned and
should not be in parallel.
The following table shows the ECG electrode label to identify each
electrode and its associated color of AHA and IEC standards.
Electrode
labels
(IEC)
Electrode
colors
(IEC)
Electrode
labels
(AHA)
Electrode
colors
(AHA)
Placement
R Red RA White
Directly below the clavicle and
near the right shoulder
L Yellow LA Black
Directly below the clavicle and
near the left shoulder
N Black RL Green On the right lower abdomen
F Green LL Red On the left lower abdomen
C1 Red V1 Red
On the fourth intercostal space
at the right sternal border
C2 Yellow V2 Yellow
On the fourth intercostal space
at the left sternal border
C3 Green V3 Green
Midway between the V2 and
V4 electrode positions
C4 Brown V4 Blue
On the fifth intercostal space at
the left midclavicular line
C5 Black V5 Orange
On the left anterior axillary
line, horizontal with the V4
electrode position
C6 Violet V6 Violet
On the left midaxillary line,
horizontal with the V4
electrode position