Patient Monitor User’s manual
7-13
PVC count indicates the quantity of PVC occurring within 1 minute. When
the PVC quantity in 1minute exceeds the set alarm limit, the monitor will
indicate alarm message “Multi PVCs”.
7.7.3 Setting Arrhythmia Alarm
Select【ARR Analysis】→【ARR Alm Setup】in the ECG setting menu,
in the popped up menu, you can set the alarm status for each Arrhythmia,
including alarm switch, alarm level and alarm print switch. You can select
【ALL ARR Alm-Resume Default】,【All ARR Alm-Alm Switch】,【ALL
ARR Alm-Alm Level】and【All ARR Alm-Alm Print】to set the alarm status
for all arrhythmia.
Arrhythmia can be analyzed by the monitor is shown in the following table:
Arrhythmia Type Applicable Patient Type
Alarm Level
ASYSTOLE All Default High
VENT FIB/TACH Non-paced Default High
MULTI PVCS Non-paced Default medium,User selectable
COUPLET Non-paced Default low,User selectable
BIGEMINY Non-paced Default low,User selectable
TRIGEMINY Non-paced Default low,User selectable
R ON T Non-paced Default low,User selectable
TACHY All Default medium,User selectable
BRADY All Default medium,User selectable
MISSED BEATS Non-paced Default low,User selectable
PNP Paced Default medium,User selectable
PNC Paced Default medium,User selectable
NOISE All Default low,User selectable
VPB Non-paced Default low,User selectable
ST HIGH
All Default low,User selectable
ST LOW
All Default low,User selectable
VTACH
Non-paced Default High
PVCS HIGH
Non-paced Default low,User selectable