Patient Monitor Users manual
9-5
the operator shall enter the SpO
2
parameter setting menu and set NIBP Same Side
toOnin order to ensure that the SpO
2
alarm status will keep constant during the
NIBP measurement until the completion of NIBP measurement. IfNIBP Same Side
is set toOff, the low perfusion will result in the SpO
2
measurement inaccuracy
during the NIBP measurement and trigger the physiological alarm of SpO
2
.
Setting QRS Volume
SelectQRS Volumein the SpO
2
setting menu to change the QRS volume. The
volume can be set to a value between 1 and 6, and also can be silenced.
Setting Average Time
The SpO
2
reading shown on the monitor is the average of data collected within a
specific time. The shorter the averaging time is, the quicker the monitor responds to
the change in the patient’s oxygen saturation level. Contrarily, the longer the
averaging time is, the slower the monitor responds to the change in the patient’s
oxygen saturation level, but the measurement accuracy will be improved. When a
critical patient is monitored, selecting shorter averaging time will help understanding
the patient’s state.
You can selectAverage Timein the SpO
2
setting menu, and select a proper time
as required.
Setting PR Source
To set the PR source, you can select the PR parameter area, then selectPR
Source. The options ofART, SpO
2
andAutoare available.
ART: The monitor takes ART as the PR source.
SpO2: The monitor takes SpO
2
as the PR source.
Auto: When the monitor can detect ART signal, the monitor will
automatically take the ART as the PR source. If failing to detect the ART
signal, the monitor will automatically take SpO
2
as the PR source.
For Masimo module:
Setting Sensitivity mode
You can set sensitivity mode of Masimo SpO
2
module according to use condition.
SelectSensitivity.in the SpO
2
setting menu with the options ofMax,Normal
orAPOD.
Max: This mode should be used for the sickest patients, where obtaining a reading
is most difficult. The mode is recommended during procedures and when clinician