Patient Monitor User’s manual
9-6
and patient contact is continuous.
【Normal】: This mode provides the best combination of sensitivity and probe-off
detection performance. The mode is recommended for the majority of patients.
【APOD】: This mode is the least sensitive in picking up a reading on patients with
low perfusion but has the best detection for probe-off conditions. The mode is useful
for patients that are at particular risk of the sensor becoming detached (pediatric,
combative, etc.)
Setting Delay Time mode
In the SpO2 setting menu, select 【Delay Time】, and you can select 【off】、【5s】、
【10s】or【15s】as required in the options.
Setting FastSat mode
The FastSat mode is clinically applicable during procedures when
detecting rapid changes in oxygen saturation is paramount such as induction,
intubation and sleep studies.
Select 【FastSat】in the SpO
2
parameter setting menu, you can select【On】or
【Off】to enable or disable the FastSat mode.
Setting SmartTone
【SmartTone】is a feature that affects pulse beep and can be selected in
the SpO
2
parameter setting menu. When you set it to【On】, it will allow the audible
pulse beep to beep when the pleth shows signs of motion. The pulse beep is
suppressed during signs of motion when SmartTone is set to【Off】.
Setting Wave Mode
You can select【Wave Mode】in the SpO
2
parameter setting menu,
and select【Resp. Out】or【Resp. In】.
View the Version of Firmware
You can select【Version】in the SpO
2
parameter setting menu,
and you can view the version of firmware.
9.5.2 Setting SpO
2
Waveform
Setting Wave Speed
Select【Wave Speed】in the SpO
2
wave setting menu, and select the wave speed