202 | Chapter A: Error Codes
Error
Code
Description Help
6031 Cannot use a 96-well plate with a
192-tube manifold.
Modify the Plate Type or Change the
Washer Manifold.
6032 Downloading Protocols is not
supported.
Service Only.
Contact BioTek TAC.
6040 Invalid baud rate Service Only.
Contact BioTek TAC.
These codes indicate an unexpected
software error that cannot be fixed
without BioTek support.
6041 Invalid data bits selection
6042 Invalid stop bits selection
6043 Invalid parity selection
6044 Serial port error Fix the COM port setting. Check the
cabling. Click the Port link and use the
drop-down menu to see all active ports.
Customize the Predefined Protocols to
avoid this error in future.
When controlling the BioStack with the
LHC, make sure the instrument's BioStack
setting is Manual.
6045 Serial write error
6046 Serial read error
6047 Checksum error Contact BioTek TAC.
6048 Serial NAK error Make sure the COM port setting is correct
and the cable is properly connected.
Restart the instrument. If error reoccurs,
contact BioTek TAC.
6049 Excess data, or not enough data,
received.
To correct these errors:
l Reset the instrument.
l Check cables, plug in only one
communication cable at a time:
USBor serial.
l Try running a different protocol.
If error reoccurs, contact BioTek TAC.
6050 Invalid message header
6051 Invalid message object
6052 Invalid message body size
6053 Serial message timeout
6054 Port handle error
6055 Read timeout value is invalid.
405™ Microplate Washer LS