405 LS-LHC Software Error Codes | 203
Error
Code
Description Help
6056 Unauthorized to open the COM port Make sure the COM port setting is correct
and the cable is properly connected.
Restart the instrument. If error reoccurs,
contact BioTek TAC.
6057 Out-of-range parameter for the open
port function.
6058 Unable to open the COM port.
6059 Unable to clear the transmission
buffer.
6060 Unable to close the port.
6061 Port is no longer available.
6062 Unhandled exception while
transmitting message
Contact BioTek TAC
6063 The selected plate type is not allowed
with this protocol step
Modify the protocol to change the plate
type.
6065 Too few data bytes received from the
instrument
Contact BioTek TAC.
6066 Ultrasonic cleaning assembly is not
installed
The protocol may have been created for a
different instrument. Make sure the
instrument settings match the current
instrument and modify the protocol.
6074 Invalid Y-axis offset value Service Only.
Contact BioTek TAC.
These codes indicate an unexpected
software error that cannot be fixed
without BioTek support.
6075 Invalid Z-axis offset value
6076 Vacuum filtration not allowed with
1536-well plates.
Modify the protocol to change the plate
type.
6087 'Move carrier home' is required when
duration exceeds 1 minute.
Contact BioTek TAC
BioTek Instruments, Inc.