Calibration Adjustment Procedure
Calibration Interval
The recommended calibration interval for the 9240 series is one year.
To enter the Calibration Menu:
Press the button, then use the soft key to select Utilities > Test/Admin > Security > Calibrate.
In the Calibration Menu use the softkeys to select a parameter from the menu to calibrate.
Voltage
Current
OVP
OCP
RTC
Figure 11.1 Calibration Parameters