8.2 Preset Weight Comparators
The Model 190 indicator has three (3) outputs, which can be configured
during setup to perform as Preset Weight Comparators “PWC”. The
output state of each PWC (before reaching the preset weight) is defined
in section 4.4 SETUP, the d out= (Digital Output) parameter.
With the PWC feature selected, during operation the indicator will
compare each enabled preset weight value with the displayed weight and
then output a signal for each preset based on the comparison results.
8.2.1 Entering Preset Weight Comparators Values
1. To enter the Preset Weight Comparators values, simultaneously
press the Fn/ key and the PRINT key.
2. The display will change to show Prt= (the prompt for Print Ticket
Format selection). Refer to section 8.1 for details on selecting a
Print Ticket Format.
3. Press the TARE key again to skip the Print Ticket Format
selection.
4. The display will change to show PSEt1=.
5. Press the TARE key to display the current value of the first
preset weight value.
6. If the setting displayed is acceptable, press the TARE key again
to save it.
7. Otherwise, use the Fn/ and UNITS/ keys to enter a new value
for the preset weight.
8. When the desired value is displayed, press the TARE key to
save it.
9. Continue with this procedure until a value for each preset to be used
has been entered.
10. After the last preset value has been entered and the TARE key
pressed, the indicator will return to the weight display.