Model 190 Installation, Technical and Operation
8400-M022-O1 Rev J
4.6 Calibration
The 190 indicator has six modes that can be used to perform calibration.
Four of the modes require a test load or test weights, one requires the
scale to be empty (and at zero) and the last uses the calibration “C”
numbers from a previous calibration.
CAL
With CAL displayed, press the TARE key. The display will change to
CAL1=. Proceed to the CAL1= parameter.
CAL?
With CAL? displayed, press the TARE key. The display will change
to no. Press the Fn/ key to toggle to yes and then press the
TARE key. The display will change to CAL1=. Proceed to the
CAL1= parameter. Otherwise, to skip Calibration, press the
TARE key to advance to the sio?. prompt.
During calibration it will be necessary to enter values using the 190
keypad.
Pressing the TARE key will cause the data entered or displayed to
be retained and the 190 to advance to the next prompt.
The functions of numeric keys are replaced by using the
UNITS/ and the Fn/ keys.
The cursor location is identified by the blinking character and can be
advanced to the left to the next position by pressing the
UNITS/ key.
Pressing the Fn/ key will change the blinking character to the next
value.