Model 190 Installation, Technical and Operation
8400-M022-O1 Rev J
4.3 Accessing Setup
4.3.1. With the 190 ON, press the Fn/ and UNITS/ key simultaneously
4.3.2. Hold both keys until the display changes to SetUP.
4.3.3. Release the keys to begin setup.
4.3.4. Press the UNITS/ key to step to the beginning point of each setup
section.
Setup Mode (starts at USA= prompt)
Analog to Digital Filtering (starts at dFLt= prompt)
Calibration (starts at CAL1= prompt)
Set Gravity Constant (starts at CALgC= prompt)
Serial Input/Output (starts at bAUd= prompt)
Print Tab Settings (starts at Port= prompt)
Display High Resolution Weight
Key Lockout Feature Setup
Configuration for Indicator Option Boards
4.3.5 If you press the TARE key at the SetUP prompt, you may
proceed through to the next section (up to and including Colors)
by pressing the TARE key.
4.3.6 To exit setup, press the Fn/ key with any of the menu selections
displayed.
NOTE: With the exception of the SEtUP prompt, the prompts
displayed for each section are different if you push the UNITS/ key
to step through the prompts instead of pressing the TARE key to
proceed through the section. For example, if you press the UNITS/
key with the SEtUP displayed, the next prompt displayed will be A-d.
If you step through the setup prompts by pressing the TARE key,
the next prompt displayed will be A-d?. In addition, at a prompt with
the ? displayed, you must press the TARE key again to proceed
with that section. To skip the section and advance you to the next
menu selection, press the TARE key twice.