Section 7: Milk System and Refrigeration Unit 3033-007B
7-14 Concordia Xpress Technical Support Manual
2900-300B
The milk weight trays are notched to ensure they are properly placed
within the on-board refrigeration unit.
Adjustment to milk level monitoring can be done by adjusting the position
of the magnet. The milk level set-point in TEST ROUTINES can be used
as a reference.
IMPORTANT: When calibrating a milk weight tray, adjust the magnet
and remove the screwdriver from the milk weight tray. The presence of
the screwdriver will affect the magnet and skew the count in TEST
ROUTINES.
Large Milk Weight Tray Assembly