3033-014B Section 14: Messages
Concordia Xpress Technical Support Manual 14-5
2900-300B
Rear milk count is 75 or
lower.
Verify milk level.
Verify milk weight tray is properly
installed and calibrated.
Right piston took too long to
complete a movement.
Check right drive components.
Right piston reference
sensor has failed or is out
of position.
Check the right piston reference
sensor for correct position and
proper operation.
Refrigeration unit current in
excess of 14.5amps.
Verify DC supply voltage to
refrigeration unit module.
Verify refrigeration unit module
operation.
Refrigeration unit current is
below 10amp.
Verify DC supply voltage to
refrigeration unit module.
Verify refrigeration unit module
operation.
Refrigeration unit door
open.
Check door for proper seal.
Refrigeration unit fan
failure.
Replace fan.
Verify fan is operating and
unobstructed, replace if necessary.
Verify ambient temperature is below
84˚F (29˚C).
See Section 15: Troubleshooting for
detailed instructions.
Short circuit on refrigeration
temperature sensor.
Navigate to CHK
TEMPERATURES >
REFRIGERATOR and verify
temperature.
Shorted sensor = 67.3°F (20°C).
Check DC board LEDs D17, D18,
D19.
Verify operation of thermal chip.
Open circuit on refrigeration
unit temperature sensor.
Navigate to CHK
TEMPERATURES >
REFRIGERATOR and verify
temperature.
Open sensor = 0.8°F (-17°C).
See Section 15: Troubleshooting for
detailed instructions.
Door to the refrigeration
unit is open.
Close refrigeration unit door.
Verify door seal is in good
condition.