5
High Speed Drill (E9010) without a bur and
appropriate bur guard in place and the col-
let locked. Damage will occur.
11. Running the shaver blade or bur without
fluid flow (dry) may cause damage to the
device.
12. Ensure the optical cassette sensor next to
the pump cassette mechanism is kept clean
at all times.
13. Do not stall handpieces, damage can occur.
14. Always inspect for bent, dull or
damaged burs, blades or drill bits
before each use. Do not attempt to
straighten or sharpen. Do not use if dam-
aged. After use, dispose of properly.
15. Always use a bur of the proper length for
the particular guard.
16. Direct contact of the rotating cutting edge
of shaver blades and burs with metallic sur-
faces and/or other hard surfaces such as
arthroscopes, cannulas, or other instruments
can cause damage to the devices. If contact
does occur, shaver blades can break, seize,
or shed metal particles. Shaver blade or bur
should be examined for damage and
replaced if necessary.
17. Do not apply excessive loading on the
shaver blade or bur. Cutting performance is
not increased with force. Excessive force or
using shaver blades or burs as a lever can
cause damage to the device including per-
manent deformation, shedding of metal
(wear), and seizure.
18. If shaver blade clogs, use ConMed Linvatec
Declogger (E9315) to declog shaver blade.
Shaver blades or burs may also be dec-
logged by removing inner assembly from
outer assembly and clearing any tissue from
the window or aspiration hole. DO NOT
disassemble Prebent Shaver Blades with
catalog number ending in “M”
or “MT” (i.e.,
C9299M and 42ULT-ZZ-MT)
.
19. United States Federal law
restricts sale of this device to or
on the order of a physician.
20. After each use, thoroughly clean the con-
troller, handpieces, attachments and acces-
sories (See “4.1 Cleaning and Sterilizing”
on page 118).
NOTES:
1. Observe all national waste management
regulations.
2. Do not dispose of Waste Electrical and
Electronic Equipment (WEEE) as unsorted
municipal waste.
Rx ONLY