3.2.4 Tubing Set Line Attachment
1. If irrigation is being used, attach the black
cassette of the tubing set to the controller
(reference “2.1.4 Tubing Set Cassette
Installation” on page 47).
2. Spike the fluid bag.
3. Prior to connecting the clear plastic
handpiece cassette of the tubing set to the
handpiece, lubricate the O-Rings with
sterile fluid.
4. While holding the nose of the handpiece
upward, place the plastic handpiece cassette
onto the post within the cassette holder
opening, perpendicular to the handpiece.
5. Twist the cassette clockwise until it is
aligned with the handpiece body. No
downward pressure is needed while turning.
6. If leakage occurs at this interface while
operating the handpiece, perform the
following to alleviate the leakage:
(a) Remove the clear plastic cassette from
the handpiece by reversing steps 4
and 5.
(b) Inspect the O-Rings to ensure they are
seated properly within the handpiece
cassette.
(c) If no problems exist, reinstall the clear
plastic cassette in the handpiece by
following steps 4 and 5.
(d) If leakage still occurs, replace the
tubing set.